The johns hopkins hospital point-of-care testing program

The Johns Hopkins Hospital Point-of-Care Testing Program Self-Study Packet
URINE DIPSTICK
Multistix® 7 and Uristix
Operator Competency
Only operators who have completed a defined training program and can demonstrate competence will be able to perform Urine Dipstick testing. Operator competency will be assessed annually by two methods. a. Successful performance of quality control (Quantimetrix Dropper Plus Level 1 and Level 2) • Having the supervisor or qualified delegate periodically observe routine work Test Order

A physician’s order, standard protocol, or order by other health professionals authorized to request
laboratory tests is required for urine dipstick testing.
Reagents and Storage
The Siemens Urine Dipstick bottles and Urine Controls are to be dated when initially opened. The “Opened Date” and “New Expiration Date” are to be recorded on the Quantimetrix Dropper Plus Urine Control bottles when stored at room temperature. Do not use the Urine Dipsticks and Urine Controls after the expiration date. 1. Multistix® 7 and Uristix urine dipsticks are stable until the manufacturer’s expiration date printed on the bottle when stored properly. Urine dipsticks are stored in the original, tightly sealed bottle at room temperature (15-30oC or 59-86oF). Do not store dipsticks in direct sunlight. Protect dipsticks from heat sources and light exposure. The Quantimetrix Dropper Plus Urine Controls are stable for one month at room temperature (18-25°C or 64-77°F). Do not store above 30°C (86°F). The Quantimetrix Dropper Plus Urine Controls can also be stored at 2-8°C (35-46°F) in a temperature monitored refrigerator. When stored refrigerated, the Urine Controls are stable until the expiration date stated on the label. Do not freeze. Copyright 2005 Johns Hopkins Hospital All rights reserved
Revision: 10/2010
The Johns Hopkins Hospital POCT Program
Self-Study Packet
Urine Dipstick- Multistix
7 and Uristix
Quality Control Label
A JHMI Quality Control label should be affixed to the bottle of urine dipsticks when it is first opened. Record the open date at the top of the QC label in the space below “Date Opened”. Quality Control Frequency
Weekly on each opened bottle of urine dipsticks.
Each time a new bottle of urine dipsticks is opened. Quality Control
Record the lot numbers and expiration dates of the urine dipsticks and controls on the Urine Dipstick QC log sheet, along with the date, time and operator initials. Ensure that the control solutions are at room temperature before use. Test the control solutions (Level 1 and Level 2) by squeezing the control bottle and drawing the control solution across all of the reagent pads, thoroughly saturating each pad. Do not aspirate excess control back into the bottle. Drain excess from dipstick. Time appropriately. Record control results on the Urine Dipstick QC Log. Record results in the units specified on the Note any QC failures and corrective action taken on the Urine Dipstick QC Log. If both levels of controls are within the acceptable ranges, record the date and operator initials on the JHMI QC label affixed to the reagent strip bottle. This bottle may be used for one week for patient testing. Quality Control – Results Documentation Summary
Record date, time, operator initials and lot numbers of dipsticks and controls on log sheet. Record the Quality Control results in the units specified on the bottle of urine dipsticks. Do not use + or – for results (example: Record blood as 3+ not +++, NEG/negative, POS/positive) Place a √ mark in the QC Pass/Fail Column to note whether QC passed or failed. Document any QC failures and corrective actions on the Urine Dipstick QC log sheet Quality Control Failures and Corrective Actions
If one or more of the urine control solutions fail to give expected results: 1. Check to make sure that the control solutions and urine dipsticks have not expired. If the repeat test fails, open a new set of urine controls and repeat. If this test fails using the new controls, open a new bottle of urine dipsticks and repeat. If the repeat test fails using the newly opened bottle of dipsticks and controls, STOP PATIENT TESTING AND REPORTING OF PATIENT RESULTS until the quality control issues are resolved and the expected results are obtained and recorded. Document all steps taken in the Comment Section of the Urine Dipstick QC Log sheet. Contact the POCT Program Office for assistance.

Copyright 2005 Johns Hopkins Hospital All rights reserved
Revision: 10/2010
The Johns Hopkins Hospital POCT Program
Self-Study Packet
Urine Dipstick- Multistix
7 and Uristix
Specimen Collection
Use at least 2 unique patient identifiers (neither to be the patient’s room number) to verify a patient’s identity when collecting samples for testing. Label the urine container (not the lid) with the patient identifiers. Samples must be fresh (not refrigerated) and tested promptly within 1 hour of collection. Gloves must be worn during sample collection and testing (Standard Precautions) Patient Testing
Check the JHMI QC label on the bottle of urine dipsticks or the Urine Dipstick QC log sheet to make sure that quality control has been performed for the week. If not, stop and perform QC before Remove a test strip from the bottle of urine dipsticks. Recap the bottle before proceeding. Gently swirl the urine container in order to thoroughly mix the urine. Immerse the pads on the dipstick in the urine sample and remove immediately. While removing the strip from the urine container, wipe the edge of the dipstick on the container Hold the dipstick level to avoid mixing reagents from adjacent pads. Hold the dipstick near the color chart on the bottle of dipsticks and read in the order and at the time intervals specified in the attached Interpretation Chart. The glucose pad is nearest the hand- held end of the dipstick. Color changes that occur after the correct reading time are
erroneous.
Record the patient’s results, along with the date, time and operator initials on the chart, EPR or When documenting the results in the patient’s chart, place a check mark (√ ) or “QC okay” next to the results to indicate that Quality control has successfully been run within one week. Results should be noted as Point-of-Care Testing results in order to distinguish them from Dispose of urine samples in toilet when testing is finished. Discard used test strips in a JHMI approved biohazard container Patient Result Reporting
Record all urine dipstick test results, including negative results, with the date, time, operator initials Record the patient results in the reporting units specified on the bottle of urine dipsticks. Do not use (+) or (–) signs to report positive and negative results. Record negative results as “Neg” or “Negative” and positive results as “Pos” or “Positive”. Record a “+++” result as 3+, a “++” result as 2+, etc. Place a check mark (√) or “QC okay” next to the patient results to indicate that external Quality Control had been successfully run within the previous week on the bottle of dipsticks used. Note the results as Point-of-Care results to differentiate results from central laboratory results. Copyright 2005 Johns Hopkins Hospital All rights reserved
Revision: 10/2010
The Johns Hopkins Hospital POCT Program
Self-Study Packet
Urine Dipstick- Multistix
7 and Uristix
Abnormal Results
Any value greater than negative for glucose, ketones, protein, blood, nitrite or leukocytes. Both the normal and abnormal urinary pH range is 5-8.5 Questionable urine dipstick results, problems with the testing procedure or results that do not match the clinical condition should be reported to the physician/provider to determine if additional follow-up is needed, i.e. repeat the point-of-care test, send a sample to the core laboratory etc. In the event of abnormal or questionable urine dipstick results, document in the patient’s record that the physician/provider has been notified. Critical Action Value
Ketones: Large amount (80-160 mg/dL)
Notify the patient’s physician/provider to determine if additional follow-up is needed. Document in the patient’s record that the physician/provider was notified and that a read-back of the Critical Action Value occurred and was verified. Copyright 2005 Johns Hopkins Hospital All rights reserved
Revision: 10/2010

The Johns Hopkins Hospital POCT Program
Self-Study Packet
Urine Dipstick- Multistix
7 and Uristix


INTERPRETATION CHART FOR URINE DIPSTICK RESULTS:

Results are read and recorded as follows: Test Interpretation
Negative, 1/10 (tr) g/dL, ¼ g/dL, ½ g/dL, 1 g/dL, 2 or more g/dL Negative, 100 mg/dL, 250 mg/dL, 500 mg/dL, 1000 mg/dL, 2000 or more mg/dL Negative, 5mg/dL, 15 mg/dL, 40 mg/dL, 80 mg/dL, 100 mg/dL Negative, Trace, 30 mg/dL, 100 mg/dL, 300 mg/dL, 2000 or more mg/dL
OR
Copyright 2005 Johns Hopkins Hospital All rights reserved
Revision: 10/2010
The Johns Hopkins Hospital POCT Program
Self-Study Packet
Urine Dipstick- Multistix
7 and Uristix
Limitations and Interferences
Substances that cause abnormal urine color, such as visible levels of blood and bilirubin and drugs containing dyes (ex. Pyridium), nitrofurantoin (ex. Macrodantin), or riboflavin may affect the readability of the reagent pad areas on the strip. The color development on the reagent pads may be masked or a color reaction produced that could be interpreted as a false positive. These specimens should be sent to the main laboratory for testing. The limitations and interferences for each analyte are as follows: Limitations and Interferences
• Test is specific for glucose. • Ketone bodies reduce the sensitivity of the test • Moderately high ketones (40 mg/dL) may cause false negative results for specimens containing small amounts of glucose (75-125 mg/dL). This combination of ketones and low glucose levels is metabolically improbable in screening. • In ketoacidosis, starvation, or with other abnormalities of carbohydrate or lipid metabolism, ketones may appear in urine at levels of 10 mg/dL or higher before serum ketones are elevated. • False Trace results may occur with highly pigmented urine specimens. • False Trace results may occur with urine containing large amounts of levodopa metabolites. • Compounds such as mesna (2-mercaptoethane sulfonic acid) that contain sulfhydryl groups may cause false positive results or an atypical color reaction. • Test is equally sensitive to myoglobin as to hemoglobin. • Capoten (captopril) may reduce the reactivity of the test. • False positive results may be the result of: a. Oxidizing contaminants (ex. Hypochlorite) b. Microbial peroxidase associated with urinary tract infections. • Bacterial growth by certain organisms in a urine specimen may cause a marked alkaline shift (pH > 8.0), usually because of urea conversion to ammonia. • Contamination of urine with skin cleansers containing chlorhexidine may affect protein results. • Test is not specific for a particular protein- proteins other than albumin can cause a positive reaction. • Test is less sensitive to mucoproteins and globulins which are generally detected at levels of 60 mg/dL or higher. • Visibly bloody urines may cause falsely elevated results. • Test is specific for nitrite. • Nitrite concentration during infection increases with the length of time the urine is retained in the bladder prior to collection. • Pink spots or pink edges should not be interpreted as a positive result. • False negative results may occur with: a. Shortened bladder incubation of the urine b. Absence of dietary nitrate. c. Presence of nonreductive pathological microbes. Leukocytes
• A strip result of Small or greater is a useful indicator of infection. • Trace results observed repeatedly may be clinically significant. • Elevated glucose concentrations (≥ 3 g/dL) may cause decreased test results. • The presence of cephalexin (Keflex), cephalothin (Keflin), or high concentrations of oxalic acid may cause decreased test results. • Tetracycline may cause decreased reactivity, and high levels of the drug may cause a false negative reaction. • Positive results may occasionally be due to contamination of the specimen by vaginal discharge. Copyright 2005 Johns Hopkins Hospital All rights reserved
Revision: 10/2010

Source: http://pathology.jhu.edu/department/staff/POCT/SelfStudy/UrineDipstick.pdf

renuderm.co.za

M e l a n o c e u t i c a l S k i n B r i g h t e n i n g s y s t e m ALL BRIGHT AND EVEN For all over face and body brightening This skin brightening and lightening face and body cream is ideal for use all over the body. Can be used on the face, neck, elbows, dark knuckles or hands, in between thighs, knee’s, etc. Contains NO irritating ingredients such as alpha hydroxy acids, therefore

huntersvillehalf.com

Top Males Overall in 5K division. Place Bib # Name Elapsed Pace _______________________________________________________ 1 740 Tony Rao 00:17:55 5:47 2 739 Jay Rao 00:18:30 5:58 Top Females Overall in 5K division. Place Bib # Name Elapsed Pace _______________________________________________________ 1 557 Sarah Hart 00:22:04 7:07 2 693 Kate Swartwout 00:22:57 7:24 Male 14 & Under in division 5K

© 2010-2017 Pdf Pills Composition