Digital Frequency Analysis of Valve Sound Phenomenain Patients after Prosthetic Valve Surgery: Its Capabilityas a True Home Monitoring of Valve FunctionDirk Fritzsche1, Thomas Eitz1, Kazutomo Minami1, Delewar Reber3, Axel Laczkovics3, UweMehlhorn4, Dieter Horstkotte2, Reiner Körfer1
Clinics for 1Thoracic and Cardiovascular Surgery and 2Cardiology, Heart and Diabetes Center North Rhine-Westphalia, BadOeynhausen, 3Clinic for Thoracic and Cardiovascular Surgery, University Bochum, 4Clinic for Cardiovascular Surgery,University Köln, GermanyBackground and aim of the study: Depending on the ed and examined meticulously. individual risk profile of a patient, disturbances of Results: Fourteen patients were found to have suspi- the functional integrity of mechanical heart valve cious signals. In 13 patients, valve dysfunction was prostheses occur in up to 2.5% of patients each year. confirmed by fluoroscopy, but in four cases neither The early phase of prosthetic dysfunction (due to transthoracic nor transesophageal echocardiography thrombus formation, tissue ingrowth or endocardi- detected abnormal occluder motion or ‘musses’ adja- tis) usually remains undiagnosed, as patients do not cent to the prosthesis. Normal valve sounds returned present with symptoms in this situation, and imag- in four patients who underwent thrombolytic thera- ing techniques (echocardiography, fluoroscopy) py. All patients regularly recorded and passed on demonstrate normal occluder motion. The delay their signals. Surveys revealed high acceptance and between the onset of prosthetic valve dysfunction easy handling of the Thrombocheck device. and its clinical manifestation may result in complica- Conclusion: Home monitoring of sound pressure tions (e.g. thromboembolism) or extended therapeu- measurements of prosthetic valves by digital fre- tic options (e.g. reoperation rather than more quency analysis via a Fast Fourier transformation intensive anticoagulation). may detect even very mild alterations of prosthetic Methods: A total of 291 patients with mechanical valve function. The next evolution of control sys- heart valves was allocated to four different sub- tems, allowing for registration of flow, frequency groups, and each measured their valve sounds regu- spectrum and electrocardiography, opens potential larly with the ‘ThromboCheck’ device. Depending applications for Internet-based, remote monitoring on the subgroup, the signals were compared with of cardiac patients. different reference signals. Patients in whom a suspi- cious signal was detected were immediately contact-
The Journal of Heart Valve Disease 2005;14:657-663
Despite major advances in the treatment of heart
probably develops over a period of several days to
valve lesions, there remains a complication rate of
weeks. Prosthetic valve dysfunction is usually discov-
between 0.8% and 6% per year for those patients who
ered only after manifestation of serious fatal events,
have undergone valve replacement surgery (1-4).
such as thromboembolic complications or heart failure.
Considering that approximately 100,000 heart valve
In order to reduce the risk of undiagnosed valve dys-
operations are performed worldwide each year, the
function between two medical consultations, a system
number of serious complications related to thrombo-
was developed which allows recording of the acoustic
sis, tissue ingrowth, leaflet tears, calcification, leakage
phenomena with minimal effort at any place world-
and prosthetic valve endocarditis still numbers several
wide, and in a manner which is economically accept-
A prosthetic, non-structural valve dysfunction which
As minimal changes in prosthetic valve function
is consistent with disturbed motion of the leaflets
were reliably detected by frequency spectra analysis
caused by thrombus formation or tissue ingrowth
(5-8), this technique has the potential to diagnose dys-functions long before hemodynamic deterioration ormanifestation of systemic complications occur.
Initially, preliminary experience was gained with
PD Dr. med. habil. Dirk Fritzsche, Abteilung für Thorax- undKardiovaskularchirurgie, Herz- und Diabeteszentrum Nordrhein-
recording the frequency spectra of the click phenome-
Westfalen, Georgstr. 11, D-32545 Bad Oeynhausen, Germany
na of prosthetic valves, their analog/digital processing
658 Home monitoring of prosthetic valves
and comparison with a reference file. The encouraging
are shown in the display of the device with a short
initial results obtained led to the introduction of a first
message (‘ok’, error > ‘repeat measurement’, 2nd error
hand-held device (the ‘ThromboCheck’) for home
monitoring of the functional state of prosthetic heartvalves. Subgroup 3
Five devices were utilized in an application trial in
Clinical material and methods
cardiological centers in Israel. In this trial, the patient’sfrequency spectrum (valve models, all bileaflet, were
This study comprised three aspects: simulation stud-
SJM, CarboMedics, Sulzer Carbomedics (Austin, TX,
ies, animal experiments, and some clinical investiga-
USA) and Sorin Bicarbon (Sorin Biomedica, Saluggia,
tions, the results of which have been published
Italy) was compared with the 90% percentile of an
previously (5-7). Herein are presented the preliminary
average frequency spectrum obtained from 50,000 pre-
results of the first multicenter application of the
clinical measurements of the frequency and amplitude
‘ThromboCheck’, and the first single-center prospec-
distribution of click sounds of different mechanical
tive, double-blind study. After appropriate training for
prosthetic valves. At present, 71 patients are involved
use of the Thrombocheck, four different sub groups of
patients were evaluated with the device. Subgroup 4 Subgroup 1 (double-blind, prospective)
Another 102 patients (68 males, 34 females; 87 with
Eighty-eight patients (51 males, 37 females; mean
AVR, 15 with MVR; sinus rhythm in 83%) provided
age 52.6 years) underwent aortic valve replacement
themselves with a ThromboCheck device late after the
(AVR) (n = 51), mitral valve replacement (MVR) (n =
implantation of a mechanical valve prosthesis (mean
30), pulmonary valve replacement (PVR) (n = 1), MVR
4.2 ± 1.7 years), having been informed of the device’s
+ tricuspid valve replacement (TVR) (n = 2), AVR +
existence via the Internet or self-help groups. The
MVR (n = 2), and PVR + TVR (n = 2). The valves
valve types included were mainly SJM and Medtronic
implanted were St. Jude Medical (SJM; St. Paul, MN,
Advantage; only two patients had received monocusp
USA) or Medtronic Advantage (Minneapolis, MN,
valves (Medtronic Hall, Minneapolis, MN, USA).
USA), both of which were bileaflet. All patients were
Apart from their prosthetic heart valves, these patients
provided with the Thrombocheck during their hospital
were healthy. Ninety-four (92%) of the patients were
stay for valve replacement surgery. Patients were
INR self-managers. As part of a voluntary interview
trained to handle the device in addition to conducting
with regard to their quality of life, the patients were
International Normalized Ratio (INR) self-manage-
prepared to provide information about the handling
ment using the ‘CoaguCheck’ device. The mean follow
and user-friendliness of the device.
up of these patients was 3.2 ± 2.1 months. The resultswere double-blinded and decoded by an independent
supervisor. Registration of the digital frequency analy-sis of the sound phenomena was encoded; this was
Among the present patient cohort, 14 patients were
transmitted every other day and decoded and con-
found to have suspicious sound signals (3.8%), and
trolled by a person unknown to either the patient or to
these were confirmed in 13 cases by using alternative
the study center. If the decoded signal demonstrated
diagnostic methods. In 13 patients a pathological valve
any suspicious sound phenomena, the study center
function was confirmed by fluoroscopy, while in four
cases neither transthoracic echocardiography (TTE)
A detailed clinical follow up focusing especially on
nor transesophageal echocardiography (TEE) revealed
potential microemboli was performed every three
a suspicious signal. A normal valve sound returned in
months by means of patient interviews.
four patients after thrombolysis had been carried out. Subgroup 2 Subgroup 1: (double-blind, prospective)
A multicenter study is currently investigating
The patients managed the handling and transfer of
whether anticoagulation therapy without coumarin
the encoded data without any problems. Prosthetic
derivatives after implantation of a mechanical valve
valve endocarditis, peripheral or central embolizations
prosthesis is reasonable. All 30 patients currently
were not diagnosed during the entire observation peri-
involved in the study were provided with a
od. Among all of the INR measurements, 87% were
ThromboCheck device and trained in its use while still
in-patients after their valve surgery. In contrast to sub-
One female patient transmitted a suspicious fre-
group 1, the results of the digital frequency analysis
quency spectrum for three consecutive days at approx-
Home monitoring of prosthetic valves 659
Figure 1: A) Frequency spectrum of a prosthetic valve withFigure 2: A) Focal point of frequency before and afterregular function. Note the characteristic opening andsurgical removal of thrombus. B) Relationship of acousticclosure sounds of a bileaflet valve. B) Frequency spectrumenergy of the two leaflets before and after surgical removalof a valve with impaired leaflet motion. Only the openingand closure sound of one leaflet is recorded.
imately six weeks postoperatively. At the same time,
two suspicious measurements, leading to a request to
the INR values shifted between 1.7 and 2.1 (INR self-
management). At one day after the INR had reachedthe target therapeutic range (≥2.8), the frequency spec-
Subgroup 3
trum was normalized. As the patient remained com-
Ten of 71 patients (11%) demonstrated a suspicious
pletely asymptomatic, no further diagnostics were
frequency analysis during routine check-up. In nine
patients an impaired valve function could be demon-
The vast majority of patients felt that the device was
strated, and one suspicious frequency analysis was
easy to use, reliable, and improved their well-being.
interpreted as a false negative. Four patients with one
Following a special adaptation of the software, the
leaflet stuck at the closed position were admitted for
frequency analysis of patients with two prosthetic
thrombolytic therapy with recombinant tissue plas-
valves was found to be reliable, even if they were not
minogen activator. At 48 h after thrombolysis, three
patients again showed an inconspicuous frequencysignal (8). Subgroup 2
In this subgroup, three patients (10%) presented with
Subgroup 4
a pathological frequency analysis. The device recorded
In this subgroup, neither clinical symptoms nor fre-
660 Home monitoring of prosthetic valvesTable I: Reported patient experiences with the
quency spectra indicative of prosthetic malfunctionwere detected during a mean observation period of 4.2
± 1.3 months. The results of the questionnaire regard-ing user-friendliness, handling, technical problems
Figure 3: A) Acoustic signal in the patient with
and quality of life are listed in Table I. undisturbed valve function shows two amplitude maximarepresenting the closure clicks of both leaflets. B) AcousticExemplary cases signal with impaired leaflet motion; only one maximum isCase no. 1
In this 64-year-old male (who underwent AVR with a
bileaflet, On-X, valve), anticoagulation therapy wasswitched from warfarin plus aspirin to aspirin alone. After three months, the ThromboCheck recorded
ic valve revealed massive thrombus formation at the
pathological frequency spectra indicative of impaired
valve prosthesis. Before and after surgery the valve
leaflet motion for three consecutive days (Fig. 1A).
sounds were recorded with a ThromboCheck. Figure
Neither TTE nor TEE identified any pathology.
2A and B illustrate the focal point of frequency and the
Fluoroscopy confirmed valve dysfunction with incom-
relationship of the acoustic energy of the two leaflets.
plete occluder motion of one leaflet. At four days after
Both parameters were changed significantly.
the first pathological frequency signal had been docu-
Subsequently, because of another operation, antico-
mented, medication with heparin plus warfarin was
agulation was switched to unfractionated heparin. At
started. After five days, the ThromboCheck signals
five days postoperatively, the ThromboCheck meas-
ured a pathological frequency spectrum typical forrestricted leaflet motion (Fig. 3A and B). Neither TTE
Case no. 2
nor TEE revealed any irregularity. X-ray fluoroscopy
This 72-year-old female (AVR bileaflet valve, SJM),
demonstrated impaired leaflet motion.
stopped oral anticoagulation prior to an orthopedicoperation. Several weeks later, the patient consulted
Case no. 3
her physician because of dyspnea. TEE revealed an
This 74-year-old male underwent AVR with a SJM
impaired motion of both leaflets, without any demon-
31-mm composite valve graft, but was admitted to
stration of thrombus. Surgical revision of the prosthet-
hospital at 38 days postoperatively for bleeding from
Home monitoring of prosthetic valves 661
esophageal varices due to liver cirrhosis. This patientwas monitored with the ThromboCheck once daily. The first signal had already documented changes with-in the primary frequency spectrum. Invasive hemody-namic monitoring with a Swan-Ganz catheter wasinconspicuous. Consecutively performed TEE andangiography revealed complete immobilization of oneleaflet, while motion of the other leaflet was regardedas ‘normal’. However, the frequency spectrum showeda significant reduction in the amplitude of both leaflets(Fig. 4A-C). Case no. 4
This 64-year-old female underwent AVR with an SJM
valve. She was admitted to hospital with suspectedprosthetic valve dysfunction; this was diagnosed byTEE, and confirmed by investigation with theThromboCheck. On reoperation, thrombotic materialwas found predominantly at the sewing ring (Fig. 5). Discussion
Previous investigations in animals and patients have
demonstrated that the acoustic sound phenomena ofprosthetic heart valves remain constant over time. However, the same prosthetic device, when implantedin various animals and patients, may produce differentsound pressures. These differences may be due to indi-
vidual flow and resonance conditions. Furthermore, itcould be demonstrated that each valve replacementdevice produces characteristic frequency spectra.
Pathological processes involving heart valve pros-
theses result in distinct changes to the acoustic signalsthat are characteristic of the respective device. Previous experimental studies have shown that thespectral analysis of click sounds indicated prostheticvalve dysfunction long before echocardiography, fluo-roscopy or hemodynamic measurements were sugges-tive of any disturbed valve function. Thrombolysisdoes not seem justified if only the frequency analysis
shows a pathological result, though the patient shouldbe intensely screened. Thus, a suitable hand-helddevice for signal recording represents an ideal option
Figure 4: A) The course of ThromboCheck measurements
for the home monitoring of prosthetic valve integrity. shows a reduction in fingerprint similarity to less than
It should be conceded that the frequency spectrum of
10%. B) The intact prosthetic valve’s frequency spectrumshows a maximum for each leaflet. The thrombosed
prosthetic clicks must be influenced by changes in
prosthetic valve’s frequency spectrum only shows a ‘single-
hemodynamics or atrial fibrillation. A screening
click’ because of complete blockage of one leaflet. The
method requires high sensitivity but not necessarily
amplitude was reduced by factor 30, representing the
high specificity, as differential diagnosis will be made
beginning of thrombus formation at the second leaflet. C)
by the cardiologist once the control device has indicat-
Fast-Fourier transformation (FFT) of the signal shows thatcertain frequencies are significantly reduced, or could no
Fourteen patients of subgroups 1, 2 and 3 were iden-
tified by pathological frequency analysis (1.1% ingroup 1; 10% in group 2; 11% in group 3). Only in onepatient in subgroup 3 was the frequency analysis inter-
662 Home monitoring of prosthetic valves
application of innovative technology in their privateliving space and healthcare. They were also askedwhether they were prepared to use telemonitoring andsmart home care in their personal surroundings. Infact, 96% of the participants supported the idea ofusing such features and systems, especially for the careand monitoring of chronically ill people. In addition,83% expected that their desire for more safety wouldbe met, and 73.3% felt that this technology wouldprobably save time and increase the mobility and inde-pendence of older patients (70.8%). Only 11.7% sug-gested that telemonitoring and smart hometechnologies were unnecessary and would only causefears (18).
Faced with this background - and considering the
fact that whilst home monitoring cannot substitute for
Figure 5: A thrombosed prosthetic valve (St. Jude Medical);
regular physician consultation, it might minimize time
thrombotic material was found predominantly at the
intervals between investigations and allow early diag-
nosis of prosthetic valve-associated complications -patients may benefit significantly from devices which
preted as a false negative. However, the subgroups
offer them greater safety and an improved quality of
included different patient populations (positive selec-
tion regarding age, compliance, INR self-management,
The efforts aimed at a maximum reduction in INR
etc.). The present results suggest that prosthetic valvu-
levels - and therefore at a further reduction in hemor-
lar dysfunction without clinically apparent symptoms
rhage-related complications - seem reasonable. It even
seems to occur more frequently than is generally pre-
seems realistic that in the near future, and under cer-
sumed. This suggestion is supported by the results of
tain conditions, a patient who has undergone heart
studies dealing with the occurrence of high-intensity
valve replacement with a mechanical prosthesis can be
treated without the use of coumarin derivatives. One
This observation could be of imminent importance,
of these conditions would be the availability of a tool
especially in developing countries where a high per-
such as the ThromboCheck, which is able to analyze
centage of the inhabitants live in rural areas and there
the frequency spectra of valve sounds. Whether in the
are huge distances between medical centers, with con-
future patients bearing mechanical prostheses may be
sequent long time intervals between INR measure-
treated solely with aspirin, for example, remains to be
ments. The ThromboCheck provides reproducible
seen. However, the pursuit of this path - which would
results of frequency analyses, and is also easy to use.
certainly lead to an elimination of hemorrhage-related
Moreover, prosthetic valve dysfunction is detected
complications - requires that clotting-related complica-
early enough to prevent severe, or even fatal, conse-
tions are detected immediately and treated when nec-
Several groups have studied the acoustic signals of
Björk-Shiley mechanical valves following strut fracture
Therapeutic considerations
(10-17), but none of the reports of in-vivo testings has
The incidence of prosthetic heart valve obstruction
presented individual baseline measurements revealing
through non-infectious mass is considered to be 0.5%
an individual, fingerprint-like frequency spectrum.
per year after AVR, and up to 2.5% per year after MVR
The present authors’ studies have contributed to the
(19). These values are considerably higher in poorly
development of the first hand-held device for home
controlled populations, for example in developing
monitoring of heart valve function, and this is current-
countries. Several studies have indicated the impact of
ly being tested under clinical conditions.
thrombolysis or valve re-replacement on prognosis
One potential disadvantage of such a device is an
(20-23). Recently, the present authors presented pre-
increase in patient concern. However, the results of the
liminary results of simulation and animal studies relat-
questionnaire indicated that patients may feel safer,
ed to the recording and transmission of acoustic sound
and that the device may help them to cope with the
phenomena (5). Further investigations have concen-
inevitable disadvantages of a mechanical heart valve.
trated on a reliable implementation of the Doppler
In a recent study, 307 persons aged 50 years or more
probe. If these efforts progress successfully, it could be
were interviewed with regard to their attitude to the
envisaged that this type of home monitoring will be
Home monitoring of prosthetic valves 663
made available to a large patient clientele. The present
DW, Stein PD. Low frequency analysis of opening
type of device will certainly be important in future for
sound for detection of single leg separation of
the early detection of relevant dysfunctions in heart
Björk-Shiley convexo-concave heart valves. J Heart
valve prostheses, as well as in the observation of med-
ically treated cardiac patients, a reduction in their anti-
11. Candy N, Jones HE. Classification of prosthetic
heart valve sounds: A parametric approach. J
thromboembolic and bleeding complications.
12. Walker DK, Scotten LN. Discrimination in vitro
References
between the acoustic emissions from Björk-Shiley
1. Horstkotte D, Piper C, Schulte HD.
convexo-concave valves with and without a broken
Thrombosierung prosthetischer Herzklappen:
minor strut. Med Biol Eng Comput 1991;29:457-464
13. Candy N, Jones HE. Processing of prosthetic heart
valve sounds for single leg separation classification.
2. Jamieson WR, Munro AI, Miyagishima RT, et al.
Multiple mechanical valve replacement surgery
14. Dow JJ, Plemons TD, Scarbrough K, et al. Acoustic
comparison of St. Jude Medical and CarboMedics
assessment of the physical integrity of Björk-Shiley
prostheses. Eur J Cardiothorac Surg 1998;13:151-159
convexo-concave heart valves. Circulation
3. Munclinger MJ, Patel JJ, Mitha AS. Thrombolysis of
thrombosed St. Jude Medical prosthetic valves:
15. Kien GA, Jeffries B, Katz HV, et al. Digital acousti-
Rethrombosis - a sign of tissue ingrowth. J Thorac
cal analysis of normal and bimodal Björk-Shiley 60°
convexo-concave heart valves. Am J Cardiol
4. Alexiou C, McDonald A, Langley SM, Dalrymple-
Hay MJ, Haw MP, Monro JL. Aortic valve replace-
16. Eberhardt A, Chassaing C, Ward M, Lewandowski
ment in children: Are mechanical prostheses a good
S. Acoustical characterization of mechanical valve
option? Eur J Cardiothorac Surg 2000;17:125-133
condition and loading. J Heart Valve Dis 1995;4:649-
5. Eitz T, Fritzsche D, Grimmig O, et al. Acoustic phe-
nomena and valve dysfunction in cardiac prosthe-
17. Plemons TD, Hovenga M. Acoustic classification of
ses: Data acquisition and collection via the internet.
the state of artificial heart valves. J Acoust Soc Am
6. Fritzsche D, Grimmig O, Eitz T, Brensing A,
18. Böhm U, Röhrig A, Schadow B. Telemonitoring and
Minami K, Koerfer R. Home monitoring of patients
smart home care. Hohe Akzeptanz bei den über 50-
after prosthetic valve replacement. Experimental
Jährigen. Deutsches Ärzteblatt 2003;50:2743-2744
background and first clinical attempts. Thorac
19. Horstkotte D, Pieper C, Schulte HD.
Thrombosierung prosthetischer Herzklappen:
7. Klein RM, Niehues R, Horstkotte D, Schulte H-D,
Diagnostik und Management. Z Kardiol 1998;87:20-
dysfunktion nach Bügelbruch von konvex-konkav
20. Shapira Y, Herz I, Birnbaum Y, Snir E, Vidne B,
Björk-Shiley Monoklappenscheibenprothesen;
Sagie A. Repeated thrombolysis in prosthetic heart
Häufigkeit, Diagnostik und Management.
valves: A report of three cases and review of the lit-
Intensivmedizin und Notfallmedizin 1995;32:462-
erature. J Heart Valve Dis 2000;9:146-149
21. Birdi I, Angelini G, Bryan A. Thrombolytic therapy
8. Ben Zekry S, Sagie A, Ben-Dor I, Nukrian H, Battler
for left sided prosthetic heart valve thrombosis. J
A, Shapira Y. Initial clinical experience with hand-
held device (Thrombocheck®) for the detection of
22. Lengyel M, Vegh G, Vandor L. Thrombolysis is
bileaflet prosthetic valve malfunction. J Heart Valve
superior to heparin for non-obstructive mitral
mechanical valve thrombosis. J Heart Valve Dis
9. Hiratsuka R, Fukunaga S, Tayama E, et al. High-
intensity transient signals due to prosthetic valve
23. Renzulli A, Vitale N, Caruso A, Dialetto G, Schinosa
obstruction: Diagnostic and therapeutic implica-
L, Cotrufo M. Thrombolysis for prosthetic valve
thrombosis: Indications and results. J Heart Valve
10. Durand LG, Grenier MC, Inderbitzen R, Wieting
COMPU-TOOLS TEEN/ADOLESCENT (ASAM) CRITERIA WORKSHEET (11/01/2004) LEVELS OF LEVEL III CRITERIA OUTPATIENT TX INTENSIVE OUTPATIENT MEDICALLY MEDICALLY DIMENSIONS MONITORED MANAGE INPATIENT INPATIENT TX Responses Responses Responses Responses Withdrawal No withdrawal risk No overt symptoms of Risk of Withdrawal but Sever
Viagra and Boniva were two standout campaigns to our marketing experts, for DTC Perspectives’ annual Creative Critique. Jonathan Isaacs of Ogilvy Healthworld and John J. Nosta of Ogilvy CommonHealth Worldwide explain how sometimes the simple and subtle approaches in marketing can create a stronger, longer-lasting impact with consumers. Following are their respective reviews. Viagra: �